Detalhes do Documento

[Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study]

Autor(es): Ferreira, Ana Isabel Leitão ; Mendes, Joaquim Gabriel ; Nunes, Catarina Sofia ; Amorim, Pedro

Data: 2019

Identificador Persistente: http://hdl.handle.net/10400.16/2441

Origem: Repositório Científico da Unidade Local de Saúde de Santo António (ULSSA)

Assunto(s): Monitor do índice bispectral; Bispectral index monitor; Profundidade da anestesia; Depth of anesthesia; Consciência; Awareness; Tempo de atraso; Time delay


Descrição

Background and objectives: According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response. Methods: Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil. Results: The difference between the predicted and the real BIS was in average 30.09±18.73s. Conclusion: Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.

Tipo de Documento Artigo científico
Idioma Português
Contribuidor(es) Repositório Científico da Unidade Local de Saúde de Santo António
Licença CC
facebook logo  linkedin logo  twitter logo 
mendeley logo

Documentos Relacionados

Não existem documentos relacionados.