A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading

Author(s): Da Cunha, Hiron Andreaza ; Francischone, Carlos Eduardo ; Nary Filho, Hugo ; Gomes De Oliveira, Rubelisa Cândido

Date: 2022

Persistent ID: http://hdl.handle.net/11449/231098

Origin: Oasisbr

Subject(s): Dental implants; Immediate loading; Placement torque; Resonance frequency


Description

Made available in DSpace on 2022-04-29T08:43:30Z (GMT). No. of bitstreams: 0 Previous issue date: 2004-01-01

Purpose: In this study, standard Brånemark System implants and Brånemark TiUnite implants were evaluated regarding primary stability and placement torque. The correlation between placement torque and primary stability as well as the influence of implant design on placement torque and primary stability were examined. Materials and Methods: Twelve patients who presented bilateral loss of either maxillary lateral incisors or premolars were treated with 24 immediately loaded dental implants. Each patient received 1 standard 3.75 x 13-mm Brånemark System implant and one 3.75 x 13-mm TiUnite Mk III implant. One at a time, each implant was connected by a transducer to an Osstell machine that automatically translated a resonance frequency value for the implant into an implant stability quotient value. Osseocare equipment was used to measure the placement torque for both types of implants. Results: Statistical analysis showed higher mean values for standard implants in relation to placement torque and resonance frequency values. Discussion: There was no overall correlation between placement torque and resonance frequency values; this finding supported previous studies. Conclusion: Stability was shown to be higher for the standard implants. Mean values of torque resistance were higher for the standard implants than for the TiUnite implants. Implant design appeared to influence primary stability and placement torque.

Dept. of Prosthodontics Specialties Section of Implantology Brazilian Dent. Assoc.-Goiás, Goiânia, Goiás

Department of Implantology Univ. of Sagrado Coracao, Bauru, São Paulo

Branemark Osseointegration Ctr., Bauru, São Paulo

Department of Operative Dentistry Faculty of Dentistry of Bauru, Bauru, São Paulo

Department of Oral Diagnostics Federal University of Goiás, Goiânia

Graduate Periodontics Univ. of State of São Paulo, Araraquara, São Paulo

Document Type Journal article
Language English
facebook logo  linkedin logo  twitter logo 
mendeley logo

Related documents

No related documents