Validation study of an automated wrist monitor, omron model HEM-608, compared with the standard methods for blood pressure measurement

Autor(es): Plavnik,Frida Liane ; Zanella,Maria Teresa

Data: 2001

Origem: Oasisbr

Assunto(s): automated blood pressure device; wrist monitor; hypertension and normotension


Descrição

OBJECTIVE - The aim of our study was to assess the profile of a wrist monitor, the Omron Model HEM-608, compared with the indirect method for blood pressure measurement. METHODS - Our study population consisted of 100 subjects, 29 being normotensive and 71 being hypertensive. Participants had their blood pressure checked 8 times with alternate techniques, 4 by the indirect method and 4 with the Omron wrist monitor. The validation criteria used to test this device were based on the internationally recognized protocols. RESULTS - Our data showed that the Omron HEM-608 reached a classification B for systolic and A for diastolic blood pressure, according to the one protocol. The mean differences between blood pressure values obtained with each of the methods were -2.3 +7.9mmHg for systolic and 0.97+5.5mmHg for diastolic blood pressure. Therefore, we considered this type of device approved according to the criteria selected. CONCLUSION - Our study leads us to conclude that this wrist monitor is not only easy to use, but also produces results very similar to those obtained by the standard indirect method.