The first implementation of a position sensitive readout for a Micro Hole and Strip Plate (MHSP) is described and tested. The readout consists on a resistive layer crossing the anodes and connected to a preamplifier on each side. By weighing the charge pulses on both preamplifiers it is possible to determine the interaction point. A 100-200 [Omega] resistance layer between consecutive strips was found to be the...
A simple system for position detection based on a Micro-Hole and Strip Plate (MHSP) using a resistive charge division method was implemented. A thin resistive layer of about 100 [Omega] between each strip (top surface) and between each anode strip (bottom surface) allows us to obtain the actual position in both orthogonal directions. The readout electronics uses only two preamplifiers, a sum and a divider circu...