Emprego de nebulizador pneumático de ICP-MS como câmara de diluição em sistemas de injeção em fluxo para determinações multielementares

Autor(es): Gomes Neto, José Anchieta [UNESP] ; Silva, José Bento B. ; Rodrigues Neto, Renato ; Curtius, Adilson José ; Souza, Ivan G.

Data: 2014

Identificador Persistente: http://hdl.handle.net/11449/25939

Origem: Oasisbr

Assunto(s): pneumatic nebulizers; ICP-MS; on-line dilution; pneumatic nebulizers; pneumatic nebulizers; ICP-MS; ICP-MS; on-line dilution; on-line dilution


Descrição

Submitted by Guilherme Lemeszenski (guilherme@nead.unesp.br) on 2013-08-22T18:48:24Z No. of bitstreams: 1 S0100-40421998000400005.pdf: 38526 bytes, checksum: 05684d0ea7e67e507273bbce1699977d (MD5)

Made available in DSpace on 2013-08-22T18:48:24Z (GMT). No. of bitstreams: 1 S0100-40421998000400005.pdf: 38526 bytes, checksum: 05684d0ea7e67e507273bbce1699977d (MD5) Previous issue date: 1998-07-01

Made available in DSpace on 2013-09-30T19:09:24Z (GMT). No. of bitstreams: 2 S0100-40421998000400005.pdf: 38526 bytes, checksum: 05684d0ea7e67e507273bbce1699977d (MD5) S0100-40421998000400005.pdf.txt: 28782 bytes, checksum: 747c4f5877a4978b6e568b99141c7456 (MD5) Previous issue date: 1998-07-01

Submitted by Vitor Silverio Rodrigues (vitorsrodrigues@reitoria.unesp.br) on 2014-05-20T14:19:43Z No. of bitstreams: 2 S0100-40421998000400005.pdf: 38526 bytes, checksum: 05684d0ea7e67e507273bbce1699977d (MD5) S0100-40421998000400005.pdf.txt: 28782 bytes, checksum: 747c4f5877a4978b6e568b99141c7456 (MD5)

Made available in DSpace on 2014-05-20T14:19:43Z (GMT). No. of bitstreams: 2 S0100-40421998000400005.pdf: 38526 bytes, checksum: 05684d0ea7e67e507273bbce1699977d (MD5) S0100-40421998000400005.pdf.txt: 28782 bytes, checksum: 747c4f5877a4978b6e568b99141c7456 (MD5) Previous issue date: 1998-07-01

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)

An automatic dispenser based on a flow-injection system used to introduce sample and analytical solution into an inductively coupled plasma mass spectrometer through a spray chamber is proposed. Analytical curves were constructed after the injection of 20 to 750 µL aliquots of a multielement standard solution (20.0 µg L-1 in Li, Be, Al, V, Cr, Mn, Ni, Co, Cu, Zn, As, Se, Sr, Ag, Cd, Ba, Tl, Pb) and the acquisition of the integrated transient signals. The linear concentration range could be extended to ca. five decades. The performance of the system was checked by analyzing a NIST 1643d reference material. Accuracy could be improved by the proper selection of the injected volume. Besides good precision (r.s.d. < 2%), the results obtained with the proposed procedure were closer to the certified values of the reference material than those obtained by direct aspiration or by injecting 125 µL of several analytical solutions and samples.

Universidade Estadual Paulista Instituto de Química Departamento de Química Analítica

Universidade Federal de Santa Catarina Departamento de Química

Universidade Estadual Paulista Instituto de Química Departamento de Química Analítica

Tipo de Documento Artigo científico
Idioma Português
facebook logo  linkedin logo  twitter logo 
mendeley logo

Documentos Relacionados