Author(s):
Silva, Guilherme Carvalho da ; Castro, Raissa Moraes de ; Forte, Marcus Davi do Nascimento ; Nogueira, Fabrício Gonzalez ; Torrico, Bismark Claure ; Lobo, Thiago Diniz ; Diniz, Denise Maria Sá Machado ; Santos, Nathiara Ellen dos ; Aguiar, Mylena Ximenes
Date: 2022
Origin: Oasisbr
Subject(s): Filtro; OOAF; TRF; Sensor de efeito hall linear; Técnicas de fisioterapia respiratória; Filtro; Filtro; OOAF; OOAF; TRF; TRF; Sensor de efeito hall linear; Sensor de efeito hall linear; Técnicas de fisioterapia respiratória; Técnicas de fisioterapia respiratória
Description
This work presents preliminary results of a digital frequency meter that has been designed to measure the frequency of oscillatory vibrations in respiratory physiotherapy devices. Maintain a frequency around 13Hz is important to keep an efficient treatment. Therefore, the proposed circuits and systems are intended to give a visual feedback to the patient with a suggestion of the optimal blow strength. A Linear Hall-Efect Sensor was coupled to the physiotherapy device and the output signal is conditioned through a set of lters. The Fast Fourier Transform of the conditioned signal is used in order to extract frequency domain information to be displayed in real-time to patients.
Este artigo apresenta os resultados preliminares do projeto de um frequencímetro digital que foi modelado para detectar as vibrações oscilatórias de um equipamente de fisioterapia respiratória. Portanto, os circuitos e sistemas propostos possuem o intuito de fornecer feedback visual para o paciente, assim como aconselhamento em relação à força desejada do sopro. Um Sensor de Efeito Hall Linear foi acoplado a um equipamento fisioterapêutico, e o sinal de saída é condicionado através de um conjunto de filtros. A Transformada Rápida de Fourier desse sinal é, então, usada para extrair informações no domínio da frequência para serem apresentadas, em tempo real, para os pacientes.