Microfluidics Integration into Low-Noise Multi-Electrode Arrays

Autor(es): Ribeiro, Mafalda ; Ali, Pamela ; Metcalfe, Benjamin ; Moschou, Despina ; Rocha, Paulo R. F.

Data: 2021

Identificador Persistente: https://hdl.handle.net/10316/105237

Origem: Estudo Geral - Universidade de Coimbra

Assunto(s): MEA; brain cells; electrical recordings; Organ-on-Chip; microfluidics; Brain-on-Chip


Descrição

Organ-on-Chip technology is commonly used as a tool to replace animal testing in drug development. Cells or tissues are cultured on a microchip to replicate organ-level functions, where measurements of the electrical activity can be taken to understand how the cell populations react to different drugs. Microfluidic structures are integrated in these devices to replicate more closely an in vivo microenvironment. Research has provided proof of principle that more accurate replications of the microenvironment result in better micro-physiological behaviour, which in turn results in a higher predictive power. This work shows a transition from a no-flow (static) multi-electrode array (MEA) to a continuous-flow (dynamic) MEA, assuring a continuous and homogeneous transfer of an electrolyte solution across the measurement chamber. The process through which the microfluidic system was designed, simulated, and fabricated is described, and electrical characterisation of the whole structure under static solution and a continuous flow rate of 80 µL/min was performed. The latter reveals minimal background disturbance, with a background noise below 30 µVpp for all flow rates and areas. This microfluidic MEA, therefore, opens new avenues for more accurate and long-term recordings in Organ-on-Chip systems.

Tipo de Documento Artigo científico
Idioma Inglês
facebook logo  linkedin logo  twitter logo 
mendeley logo

Documentos Relacionados

Não existem documentos relacionados.