Detalhes do Documento

Construction and evaluation of cysteine selective electrodes for FIA analysis of pharmaceuticals

Autor(es): Ferreira Sales, Maria Goreti ; Pille, Annouschka ; Paíga, Paula

Data: 2003

Identificador Persistente: http://hdl.handle.net/10400.22/3000

Origem: Repositório Científico do Instituto Politécnico do Porto

Assunto(s): Cysteine; Potentiometry; Ion-selective electrode; FIA and pharmaceuticals


Descrição

A flow injection analysis (FIA) system comprising a cysteine selective electrode as detection system was developed for determination of this amino acid in pharmaceuticals. Several electrodes were constructed for this purpose, having PVC membranes with different ionic exchangers and mediator solvents. Better working characteristics were attained with membranes comprising o-nitrophenyl octyl ether as mediator solvent and a tetraphenylborate based ionic-sensor. Injection of 500 µL standard solutions into an ionic strength adjuster carrier (3x10-3 M) of barium chloride flowing at 2.4mL min-1, showed linearity ranges from 5.0x10-5 to 5.0x10-3 M, with slopes of 76.4±0.6mV decade-1 and R2>0.9935. Slope decreased significantly under the requirement of a pH adjustment, selected at 4.5. Interference of several compounds (sodium, potassium, magnesium, barium, glucose, fructose, and sucrose) was estimated by potentiometric selectivity coefficients and considered negligible. Analysis of real samples were performed and considered accurate, with a relative error to an independent method of +2.7%.

Tipo de Documento Artigo científico
Idioma Inglês
Contribuidor(es) REPOSITÓRIO P.PORTO
facebook logo  linkedin logo  twitter logo 
mendeley logo

Documentos Relacionados

Não existem documentos relacionados.