Multi-slice Energy Resolving CT system using a THCOBRA detector operating in NeCH4

Ver documento

Detalhes bibliográficos
Resumo:A 2D-Thick-COBRA based detector with 10 × 10 cm2 active area operating in Ne/CH4 (95/5) flow mode, was applied to X-ray Computed Tomography (CT). The CT system is based on a third generation scanner with the object motion provided by a stepper motor. Multi-slice acquisitions of a Poly(methyl methacrylate) phantom and of a sea snail were done and the cross-sectional images were used to perform tri-dimensional (3D) visualization of the objects. Both kind of images (cross-sectional and 3D) are presented. Cross-sectional images using energy intervals of the spectrum are shown. Also, a simulation study of the influence of discharges (intrinsic to the MicroPatterned Gaseous Detector operation) in the resulting cross-sectional image was done and presented, showing no significant influence on the final image
Autores principais:Carramate, L.F.N.D.
Outros Autores:Silva, A.L.M.; Azevedo, C.D.R.; De Francesco, S.; da Silva, A.M.; Veloso, J.F.C.A.
Assunto:X-ray detectors Thick-COBRA Micropattern gaseous detectors Computerized tomography(CT)
Ano:2019
País:Portugal
Tipo de documento:artigo
Tipo de acesso:acesso aberto
Instituição associada:Universidade de Aveiro
Idioma:inglês
Origem:RIA - Repositório Institucional da Universidade de Aveiro
Descrição
Resumo:A 2D-Thick-COBRA based detector with 10 × 10 cm2 active area operating in Ne/CH4 (95/5) flow mode, was applied to X-ray Computed Tomography (CT). The CT system is based on a third generation scanner with the object motion provided by a stepper motor. Multi-slice acquisitions of a Poly(methyl methacrylate) phantom and of a sea snail were done and the cross-sectional images were used to perform tri-dimensional (3D) visualization of the objects. Both kind of images (cross-sectional and 3D) are presented. Cross-sectional images using energy intervals of the spectrum are shown. Also, a simulation study of the influence of discharges (intrinsic to the MicroPatterned Gaseous Detector operation) in the resulting cross-sectional image was done and presented, showing no significant influence on the final image