Publicação
Paediatric phantom dose optimization using digital radiography with variation of exposure parameters and filtration whilst minimising image quality impairment
| Resumo: | Objective: To induce a reduction in dose, using a paediatric phantom, through the variation of exposure parameters and filtration, without adversely affecting image quality. Methods: All images were acquired using a Kyoto Kagaku paediatric phantom and a Canon DR detector. The phantom was positioned supine for all projections: wrist (DP, lateral) and ribs (AP, oblique). Three dose protocols were established using different mAs values (high, medium and low) and copper (Cu) filtration was added to each protocol. DAP was used to calculate the ESD for each exposure. Using ImageJ, CNR was calculated for the physical measurement of image quality. Image quality was assessed by fifteen observers (visual grading analysis (VGA)). Results: The highest doses were recorded with the high dose protocol, ranging from 5.60-39.22μGy for the wrist and 5.33-129.67μGy for the ribs. When increasing the Cu filtration a decrease in ESD was observed. A difference of 0.1 in VGA score was noted between high and low dose protocols without the use of filtration, while a difference of 0.3 was noted when using filtration. As mAs increased, VGA scores increased. Fracture visibility was minimally affected by Cu filtration or projection variation. Conclusion: The variation of exposure parameters in digital radiography can achieve a dose reduction without impairing image quality in bone fractures. Superior image quality can be achieved for DP and lateral wrist projections without Cu filtration. However, the addition of Cu filtration for the rib projections has almost no impact on overall image quality. |
|---|---|
| Autores principais: | Lança, Luís |
| Outros Autores: | Bowdler, M. W.; Creedon, J.; Dayer, V.; Stensholt, N.; Pinhão, S.; Visser, M. B.; Jorge, José |
| Assunto: | Radiology Digital radiography Dosimetry Phantom Children Exposure parameters Filtration |
| Ano: | 2017 |
| País: | Portugal |
| Tipo de documento: | capítulo de livro |
| Tipo de acesso: | acesso aberto |
| Instituição associada: | Instituto Politécnico de Lisboa |
| Idioma: | inglês |
| Origem: | Repositório Científico do Instituto Politécnico de Lisboa |
Registos relacionados
groups Paediatric phantom dose study using digital radiography with variation of exposure parameters and filtration
por: Lança, Luís
Publicado em: (2018)
por: Lança, Luís
Publicado em: (2018)
groups Lesion detection performance in an anthropomorphic chest phantom: comparative analysis of low-dose CT data on two hybrid imaging systems
por: Jessop, M.
Publicado em: (2014)
por: Jessop, M.
Publicado em: (2014)
article Evaluation of exposure index (IgM) in orthopaedic radiography
por: Lança, Luís
Publicado em: (2008)
por: Lança, Luís
Publicado em: (2008)
book Radiography: impact of lower tube voltages on image quality and radiation dose in chest phantom radiography for averaged sized and larger patients
por: Hauge, I. H.
Publicado em: (2017)
por: Hauge, I. H.
Publicado em: (2017)
book Analysis of image quality and effective dose in adult chest phantom radiography with high BMI
por: Sanderud, Audun
Publicado em: (2017)
por: Sanderud, Audun
Publicado em: (2017)
book Reducing effective dose to a paediatric phantom by using different combinations of kVp, mAs and additional filtration whilst maintaining image quality
por: Bloomfield, Charlotte
Publicado em: (2015)
por: Bloomfield, Charlotte
Publicado em: (2015)
book Optimisation of exposure parameters for spinal curvature measurements in paediatric radiography
por: Reis, Cláudia
Publicado em: (2015)
por: Reis, Cláudia
Publicado em: (2015)
groups 10 kVp rule - An anthropomorphic pelvis phantom imaging study using a CR system: impact on image quality and effective dose using AEC and manual mode
por: Lança, Luís
Publicado em: (2014)
por: Lança, Luís
Publicado em: (2014)
groups 10 kVp rule – An anthropomorphic pelvis phantom imaging study using a CR system: impact on image quality and effective dose using AEC and manual mode
por: Lança, Luís
Publicado em: (2013)
por: Lança, Luís
Publicado em: (2013)
article Digital radiography detectors - A technical overview: Part 1
por: Lança, Luís
Publicado em: (2009)
por: Lança, Luís
Publicado em: (2009)
article Digital radiography detectors - A technical overview: Part 2
por: Lança, Luís
Publicado em: (2009)
por: Lança, Luís
Publicado em: (2009)
book X Radiation dose implications in screening patients with ferromagnetic IOFBs prior to MRI: a literary review
por: Jessop, Sarah
Publicado em: (2015)
por: Jessop, Sarah
Publicado em: (2015)
book Are physical measures good indicators of image quality at low dose levels? A pilot study
por: Lança, Luís
Publicado em: (2016)
por: Lança, Luís
Publicado em: (2016)
groups Effect of technical parameters on dose and image quality in a computed radiography system
por: Tavares, A.
Publicado em: (2015)
por: Tavares, A.
Publicado em: (2015)
book Image enhancement for digital radiography
por: Lança, Luís
Publicado em: (2013)
por: Lança, Luís
Publicado em: (2013)
article A multi institutional comparison of imaging dose and technique protocols for neonatal chest radiography
por: Gunn, C.
Publicado em: (2020)
por: Gunn, C.
Publicado em: (2020)
groups Are physical measures good indicators of clinical image quality at low dose levels? A pilot study
por: Lança, Luís
Publicado em: (2016)
por: Lança, Luís
Publicado em: (2016)
school Radiological imaging in digital systems: the effect of exposure parameters in diagnostic quality and patient dose
por: Lança, Luís
Publicado em: (2011)
por: Lança, Luís
Publicado em: (2011)
article Curricula, attributes and clinical experiences of radiography programs in four European educational institutions
por: Reis, Cláudia
Publicado em: (2018)
por: Reis, Cláudia
Publicado em: (2018)
book Optimisation of paediatrics computed radiography for full spine curvature measurements using a phantom: a pilot study
por: Reis, Cláudia
Publicado em: (2015)
por: Reis, Cláudia
Publicado em: (2015)
book Digital imaging systems for plain radiography
por: Lança, Luís
Publicado em: (2013)
por: Lança, Luís
Publicado em: (2013)
groups Dose and image quality in computed tomography: an overview of Portuguese data
por: Coutinho, Ana Margarida
Publicado em: (2017)
por: Coutinho, Ana Margarida
Publicado em: (2017)
school Cálculos dosimétricos em exames de tomografia computorizada ao crânio
por: Mano, Susana Margarida Varela Marques da Costa
Publicado em: (2018)
por: Mano, Susana Margarida Varela Marques da Costa
Publicado em: (2018)
article A comparative study about motivations, expectations and future plans for professional development in four European radiography programs
por: Reis, Cláudia
Publicado em: (2018)
por: Reis, Cláudia
Publicado em: (2018)
book Digital radiography detectors: a technical overview
por: Lança, Luís
Publicado em: (2013)
por: Lança, Luís
Publicado em: (2013)
article Evaluation of exposure parameters in plain radiography: a comparative study with european guidelines
por: Lança, Luís
Publicado em: (2008)
por: Lança, Luís
Publicado em: (2008)
groups Image quality and dose analysis in PA chest x-ray: comparison between AEC and manual mode acquisition using the 10 kVp rule
por: Reis, Cláudia
Publicado em: (2014)
por: Reis, Cláudia
Publicado em: (2014)
book Aneurysm clips in brain imaging using CBCT: the development of a phantom and the influence of kvp and metal suppression on image quality
por: Harsaker, V.
Publicado em: (2017)
por: Harsaker, V.
Publicado em: (2017)
article Guest editorial: OPTIMAX 2013
por: Lança, Luís
Publicado em: (2014)
por: Lança, Luís
Publicado em: (2014)
book Digital radiography detector performance
por: Lança, Luís
Publicado em: (2013)
por: Lança, Luís
Publicado em: (2013)
mic Potencial de redução de dose em tomografia computorizada com técnicas de reconstrução iterativa
por: Lança, Luís
Publicado em: (2015)
por: Lança, Luís
Publicado em: (2015)
school Efeito dos parâmetros técnicos na dose e qualidade de imagem num sistema de radiografia computorizada
por: Tavares, Amilton Joel Pereira
Publicado em: (2013)
por: Tavares, Amilton Joel Pereira
Publicado em: (2013)
article A novel method for comparing radiation dose and image quality, between and within different x-ray units in a series of hospitals
por: Al-Murshedi, Sadeq
Publicado em: (2018)
por: Al-Murshedi, Sadeq
Publicado em: (2018)
groups Assessment of scatter radiation in the shielding of radiography procedures
por: Duarte, Maria
Publicado em: (2022)
por: Duarte, Maria
Publicado em: (2022)
article Assessment of the mean glandular dose using LiF:Mg,Ti, LiF:Mg,Cu,P, Li2B4O7:Mn and Li2B4O7:Cu TL detectors in mammography radiation fields
por: Fartaria, M. J.
Publicado em: (2016)
por: Fartaria, M. J.
Publicado em: (2016)
school Filtration efficiency of meltblown webs
por: Pereira, Diana Filipa Fernandes
Publicado em: (2016)
por: Pereira, Diana Filipa Fernandes
Publicado em: (2016)
article Increasing source to image distance for AP pelvis imaging: impact on radiation dose and image quality
por: Tugwell, J.
Publicado em: (2014)
por: Tugwell, J.
Publicado em: (2014)
book The impact of pitch values on image quality and radiation dose in an abdominal adult phantom using CT
por: Lança, Luís
Publicado em: (2016)
por: Lança, Luís
Publicado em: (2016)
book Assessment of patient dose in digital systems
por: Lança, Luís
Publicado em: (2013)
por: Lança, Luís
Publicado em: (2013)
school The phantom of the Virgin
por: Pereyra Sarmiento, Patricia Susana
Publicado em: (2024)
por: Pereyra Sarmiento, Patricia Susana
Publicado em: (2024)
Registos relacionados
-
groups Paediatric phantom dose study using digital radiography with variation of exposure parameters and filtration
por: Lança, Luís
Publicado em: (2018) -
groups Lesion detection performance in an anthropomorphic chest phantom: comparative analysis of low-dose CT data on two hybrid imaging systems
por: Jessop, M.
Publicado em: (2014) -
article Evaluation of exposure index (IgM) in orthopaedic radiography
por: Lança, Luís
Publicado em: (2008) -
book Radiography: impact of lower tube voltages on image quality and radiation dose in chest phantom radiography for averaged sized and larger patients
por: Hauge, I. H.
Publicado em: (2017) -
book Analysis of image quality and effective dose in adult chest phantom radiography with high BMI
por: Sanderud, Audun
Publicado em: (2017)