Publicação

A review of squeaking in ceramic total hip prostheses

Ver documento

Detalhes bibliográficos
Resumo:The occurrence of audible squeaking in some patients with ceramic-on-ceramic (CoC) hip prostheses is a cause for concern. Considering multifactor contributing to this phenomenon, many studies have been conducted over the last decade. Great efforts have been put on understanding the mechanics of the hip squeaking to gain a deep insight into factors resulting in sound emission from hip articulation. Disruption of fluid-film lubrication and friction were reported as main potential causes of hip squeaking, while patient and surgical factors as well as design and material of hip implants were identified as affecting factors. This review article therefore summarised the recent available literature on this subject to provide a platform for future developments. Moreover, high wear rates and ceramic liner fracture as viable consequences of hip squeaking were discussed.
Autores principais:Askari, Ehsan
Outros Autores:Flores, Paulo; Dabirrahmani, Danè; Appleyard, Richard
Assunto:Ceramic total hip arthroplasty Squeaking Fluid-film lubrication Friction Multibody dynamics methodology Complexeigenvaluemethod
Ano:2016
País:Portugal
Tipo de documento:artigo
Tipo de acesso:acesso aberto
Instituição associada:Universidade do Minho
Idioma:inglês
Origem:RepositóriUM - Universidade do Minho

Registos relacionados