Resumo:Theoretical analyses show that in a sampled dc polarography on the static mercury drop electrode the response may consist of a pre-wave and a main wave that are separated by a minimum. This is a general characteristic of electrode reactions complicated by strong adsorption of the product. A steep pre-wave and the minimum are the indications of Frumkin isotherm. A flat plateau of the pre-wave is a special case of the steep plateau which appears at lower reactant concentrations. If the adsorption is weak, the pre-wave and the main wave are usually merged into a single wave.
Autores principais:Komorsky-Lovri,Šebojka
Outros Autores:Lovri,Milivoj
Assunto:polarography adsorption mercury electrode
Ano:2006
País:Portugal
Tipo de documento:artigo
Tipo de acesso:acesso aberto
Instituição associada:Fundação para a Ciência e Tecnologia
Idioma:inglês
Origem:SciELO Portugal
Descrição
Resumo:Theoretical analyses show that in a sampled dc polarography on the static mercury drop electrode the response may consist of a pre-wave and a main wave that are separated by a minimum. This is a general characteristic of electrode reactions complicated by strong adsorption of the product. A steep pre-wave and the minimum are the indications of Frumkin isotherm. A flat plateau of the pre-wave is a special case of the steep plateau which appears at lower reactant concentrations. If the adsorption is weak, the pre-wave and the main wave are usually merged into a single wave.