Publicação
One dimensional consolidation and direct shear tests: setup based on a labview approach
| Resumo: | This work describes an experimental setup that was developed in order to automate the one-dimensional consolidation and the direct shear Tests. This experimental setup assures repeatability in the data acquisition, avoiding human errors, mainly when the tests data vary with a high dynamic. The described setup is based on LabVIEW, LVDT sensors and a 16 Bit Data Acquisition Board. For the one-dimensional consolidation test it was used a Load device and a consolidometer, being the experimental setup developed according to the standard ASTM D2435 / D2435M - 11. For the direct shear Test it was used an apparatus, covered in ASTM standard D-3080 / D3080M - 11, ”Standard Method for direct shear test on soils under consolidated drained conditions”. |
|---|---|
| Autores principais: | Gonçalves, José |
| Outros Autores: | Batista, José; Paula, António Miguel; Braz-César, Manuel |
| Assunto: | Direct shear test LabVIEW One dimensional consolidation test |
| Ano: | 2019 |
| País: | Portugal |
| Tipo de documento: | comunicação em conferência |
| Tipo de acesso: | acesso restrito |
| Instituição associada: | Instituto Politécnico de Bragança |
| Idioma: | inglês |
| Origem: | Biblioteca Digital do IPB |
| Resumo: | This work describes an experimental setup that was developed in order to automate the one-dimensional consolidation and the direct shear Tests. This experimental setup assures repeatability in the data acquisition, avoiding human errors, mainly when the tests data vary with a high dynamic. The described setup is based on LabVIEW, LVDT sensors and a 16 Bit Data Acquisition Board. For the one-dimensional consolidation test it was used a Load device and a consolidometer, being the experimental setup developed according to the standard ASTM D2435 / D2435M - 11. For the direct shear Test it was used an apparatus, covered in ASTM standard D-3080 / D3080M - 11, ”Standard Method for direct shear test on soils under consolidated drained conditions”. |
|---|