Resumo:Use error is one of the leading causes of medical device incidents. It is crucial for all stakeholders to have a unified means to better understand, classify, communicate, and prevent/avoid medical device use errors. In this paper, we present our ongoing work on developing a new use error taxonomy for medical devices that has the potential to enable fine-grained analysis of use errors and their root causes in system design. Our ultimate goal is to create a generic framework that can be used by medical device designers to better identify effective design solutions to mitigating use errors.
Autores principais:Silva, Carlos
Outros Autores:Masci, Paolo; Zhang, Yi; Jones, Paul; Campos, José C.
Assunto:Human-machine interface (HMI) Medical devices Use error
Ano:2019
País:Portugal
Tipo de documento:artigo
Tipo de acesso:acesso aberto
Instituição associada:Universidade do Minho
Idioma:inglês
Origem:RepositóriUM - Universidade do Minho
Descrição
Resumo:Use error is one of the leading causes of medical device incidents. It is crucial for all stakeholders to have a unified means to better understand, classify, communicate, and prevent/avoid medical device use errors. In this paper, we present our ongoing work on developing a new use error taxonomy for medical devices that has the potential to enable fine-grained analysis of use errors and their root causes in system design. Our ultimate goal is to create a generic framework that can be used by medical device designers to better identify effective design solutions to mitigating use errors.